QuickSplit™
Post-Column Fixed Flow Splitters

- Post-column splitting is fairly straight forward. Like pre-column
flow splitting, any significant additional pressure (resistance)
down stream from the splitter may affect the split ratio. Post-column
devices also contribute to chromatographic dispersion so care must
given to connecting tubing and fittings, especially at low flow
rates. The pressure drop specification for both analytical
and semipreparative fixed splitters is 500 PSI with water at a flow
rate of 1.0 mL/min. and 20.0 mL/min. respectively. Splitters
are shipped complete with fluid resistors installed. Higher inlet
flow rates may result in a larger pressure drop across the flow
splitter. To assure a maximum pressure drop of 500 PSI across the
splitter, please specify the inlet flow when the splitter is ordered.
Custom Split Ratios
- Split ratios and resistor cartridges other than those listed
below can be ordered from
ASI
to custom configure the
QuickSplit Adjustable
Flow Splitter. Please contact technical support at 800-344-4340
for additional information about custom splitters. We will gladly
assist you in determining the best splitter configuration for your
application.
QuickSplit™ Binary Fixed Flow Splitters
-
Analytical Splitters - Binary
Analytical range, 0.1mL/min. to 5 mL/min. input flow
- These splitters will produce under 500 PSI backpressure with water at 1 .0 mL/min.* The backpressure will decrease or increase in proportion to flow rate changes. In order to assure <500 PSI pressure drop across the splitter, please
specify the inlet flow when you order.
| Description |
|
|
ASI Part Number |
| Analytical Fixed Flow Splitter, Post-Column |
Split Ratio = |
2,000:1 |
620-PO1 0-03 |
| Analytical Fixed Flow Splitter, Post-Column |
Split Ratio = |
1,000:1 |
620-PO1 0-04 |
| Analytical Fixed Flow Splitter, Post-Column |
Split Ratio = |
500:1 |
620-PO1 0-05 |
| Analytical Fixed Flow Splitter, Post-Column |
Split Ratio = |
200:1 |
620-PO1 0-06 |
| Analytical Fixed Flow Splitter, Post-Column |
Split Ratio = |
100:1 |
620-PO1 0-07 |
| Analytical Fixed Flow Splitter, Post-Column |
Split Ratio = |
50:1 |
620-PO1 0-08 |
| Analytical Fixed Flow Splitter, Post-Column |
Split Ratio = |
20:1 |
620-PO1 0-09 |
| Analytical Fixed Flow Splitter, Post-Column |
Split Ratio = |
10:1 |
620-PO1 0-10 |
| Analytical Fixed Flow Splitter, Post-Column |
Split Ratio = |
5:1 |
620-PO1 0-1 1 |
| Analytical Fixed Flow Splitter, Post-Column |
Split Ratio = |
3:1 |
620-PO1 0-12 |
| Analytical Fixed Flow Splitter, Post-Column |
Split Ratio = |
1:1 |
620-PO1 0-13 |
| Analytical Fixed Flow Splitter, Post-Column |
Split Ratio = |
Custom |
620-PO1 0-CS |
-
Analytical Replacement Resistor Sets - Binary
| Description |
|
|
ASI Part Number |
| Analytical Fixed Flow Splitter Resistor Set, Post-Column |
Split Ratio = |
2,000:1 |
620-1110-03 |
| Analytical Fixed Flow Splitter Resistor Set, Post-Column |
Split Ratio = |
1,000:1 |
620-1110-04 |
| Analytical Fixed Flow Splitter Resistor Set, Post-Column |
Split Ratio = |
500:1 |
620-1110-05 |
| Analytical Fixed Flow Splitter Resistor Set, Post-Column |
Split Ratio = |
200:1 |
620-1110-06 |
| Analytical Fixed Flow Splitter Resistor Set, Post-Column |
Split Ratio = |
100:1 |
620-1110-07 |
| Analytical Fixed Flow Splitter Resistor Set, Post-Column |
Split Ratio = |
50:1 |
620-1110-08 |
| Analytical Fixed Flow Splitter Resistor Set, Post-Column |
Split Ratio = |
20:1 |
620-1110-09 |
| Analytical Fixed Flow Splitter Resistor Set, Post-Column |
Split Ratio = |
10:1 |
620-1110-10 |
| Analytical Fixed Flow Splitter Resistor Set, Post-Column |
Split Ratio = |
5:1 |
620-1110-11 |
| Analytical Fixed Flow Splitter Resistor Set, Post-Column |
Split Ratio = |
3:1 |
620-1110-12 |
| Analytical Fixed Flow Splitter Resistor Set, Post-Column |
Split Ratio = |
1:1 |
620-1110-13 |
| Analytical Fixed Flow Splitter Resistor Set, Post-Column |
Split Ratio = |
Custom |
620-1110-CS |
-
Semi-Preparative Splitters - Binary
Semi-Prep range, 5mL/min. to 40 mL/min. input flow
- These splitters will produce under 500 PSI backpressure with
water at 20.0 mL/min. The backpressure will decrease in direct proportion
to a decrease in flow rate. Flow rates higher than 20.0 mL/min.
can be used, but may result in a higher pressure drop across the
splitter. In order to assure <500 PSI pressure drop across the
splitter, please specify the inlet flow when you order.
| Description |
|
|
ASI Part Number |
| Semi-Preparative Fixed Flow Splitter, Post-Column |
Split Ratio = |
20,000:1 |
620-PO20-00 |
| Semi-Preparative Fixed Flow Splitter, Post-Column |
Split Ratio = |
10,000:1 |
620-PO20-01 |
| Semi-Preparative Fixed Flow Splitter, Post-Column |
Split Ratio = |
5,000:1 |
620-PO20-02 |
| Semi-Preparative Fixed Flow Splitter, Post-Column |
Split Ratio = |
2,000:1 |
620-PO20-03 |
| Semi-Preparative Fixed Flow Splitter, Post-Column |
Split Ratio = |
1,000:1 |
620-PO20-04 |
| Semi-Preparative Fixed Flow Splitter, Post-Column |
Split Ratio = |
500:1 |
620-PO20-05 |
| Semi-Preparative Fixed Flow Splitter, Post-Column |
Split Ratio = |
Custom |
620-PO20-CS |
| Does not come with mounting bracket. Mounting bracket is not available. |
-
Semi-Preparative Replacement Resistor Cartridges
- Binary
| Description |
|
|
ASI Part Number |
Semi-Preparative Fixed Flow Splitter
High Flow Resistor Cartridge, Post-Column |
Split Ratio = |
20,000:1 |
620-11 20-00 H |
Semi-Preparative Fixed Flow Splitter
High Flow Resistor Cartridge, Post-Column |
Split Ratio = |
10,000:1 |
620-1120-01 H |
Semi-Preparative Fixed Flow Splitter
High Flow Resistor Cartridge, Post-Column |
Split Ratio = |
5,000:1 |
620-11 20-02 H |
Semi-Preparative Fixed Flow Splitter
High Flow Resistor Cartridge, Post-Column |
Split Ratio = |
2,000:1 |
620-1 120-03H |
Semi-Preparative Fixed Flow Splitter
High Flow Resistor Cartridge, Post-Column |
Split Ratio = |
1,000:1 |
620-1 120-04H |
Semi-Preparative Fixed Flow Splitter
High Flow Resistor Cartridge, Post-Column |
Split Ratio = |
500:1 |
620-1 120-05H |
Semi-Preparative Fixed Flow Splitter
High Flow Resistor Cartridge, Post-Column |
Split Ratio = |
Custom |
620-11 20-CSH |
-
Preparative Splitters - Binary
Prep range, 30 mL/min. to 200 mL/min. input flow
-
These splitters will produce under 500 PSI backpressure with water at the specified inlet flow rate.* The backpressure will decrease or increase in proportion to flow rate changes. In order to assure <500 PSI pressure drop across the
splitter, please contact ASI for assistance in configuring your Fixed Splitter prior to ordering.
| Description |
|
|
ASI Part Number |
| Preparative Fixed Flow Splitter, Post-Column Input Flow Range: 30 - 200 mL/min. |
Split Ratio = |
Custom |
620-PO40-CS |
Does not come with mounting bracket. Mounting bracket is not available.
-
Preparative Replacement Resistor Cartridge/Resistor Set - Binary
| Description |
|
|
ASI Part Number |
|
Preparative Fixed Flow Splitter, Post-Column, High Flow Resistor Cartridge
Input Flow Range: 30 - 200 mL/min. |
Split Ratio = |
Custom |
620-11 40-CSH |
-
High-Preparative Splitters - Binary
High-Prep range, 200 mL/min. to 1,000 mL/min. input flow
-
These splitters will produce under 500 PSI backpressure with water at the specified inlet flow rate.* The backpressure will decrease or increase in proportion to flow rate changes. In order to assure <500 PSI pressure drop across the
splitter, please contact ASI for assistance in configuring your Fixed Splitter prior to ordering.
| Description |
|
|
ASI Part Number |
| High-Preparative Fixed Flow Splitter, Post-Column Input Flow Range: 200 - 1,000 mL/min. |
Split Ratio = |
Custom |
620-PO60-CS |
Does not come with mounting bracket. Mounting bracket order information.
High-Preparative Replacement Resistor Cartridge/Resistor Set - Binary
| Description |
|
|
ASI Part Number |
High-Preparative Fixed Flow Splitter, Post-Column, Resistor Set
Input Flow Range: 200 - 1,000 mL/min. |
Split Ratio = |
Custom |
620-1160-CS |
* Flow rates higher than the specified inlet flow rate can be used, but may result in a higher pressure drop across the splitter.
|
|