Low Flow Series cartridges: 50 µL through 250 µL are interchangeable and fit in all Low Flow Series In-line, Binary tee and Ternary tee housings.
Delay volume specified below was calculated based on an ASI engineering drawing. The delay volume can be verified by injecting a sample at a known flow rate, then measuring the time this sample first appears at the exit of the mixer.
Back pressure varies directly with flow rate. For example, if the back pressure is 2 PSI at 1 mL/min., it will be .5 PSI at 250 µL/min.
High Temperature rating above 40°C is available.
If you need the mounting tap holes at the bottom, you will need to buy a housing separately. The mounting option is available only for Binary housing.
Port size: 10-32 CPI (standard HPLC fittings for 1/16" tube)
Compatible with Thermo Scientific™ nanoViper™ Fingertight Fittings
Maximum Operating Pressure: 6,000 PSI
Maximum Temperature Rating: 40 °C
Delay volume: 100 µL
Back pressure: < 2 PSI @ 1 mL/min. with water