Our default prep inlet flow for calibration is 100 mL/min. In order to assure optimum pressure drop across the splitter, please specify the prep inlet flow if it is significantly different from the default calibration prep inlet flow rate. You can specify the prep inlet flow (mL/min.) within a range of 50 - 150 mL/min. under Custom Specification field.

 

Note:

To check if the low flow capillary resistor is clogged, you can disconnect the capillary resistor from the make up inlet tee, run the splitter at approximately 1,000 PSI, and look for flow out of loose end of capillary.  You will need to give it some time to accumulate.  If the low flow capillary resistor is clogged, you will need to order Adjustable Makeup-Flow Splitter, Low Flow Capillary Resistor, ASI P/N: 690-POCR https://www.hplc-asi.com/adjustable-makeup-flow-splitter-low-flow-capillary-resistor/.

Current stock: 0

50 - 150 mL/min. Prep Inlet Flow, Adjustable Makeup-Flow Splitter

US$3,678.00
Part Number: 691-PO50-05VF
Split Flow *
Specifications:

Port size: 10-32 CPI (standard HPLC fittings for 1/16" tube)

Compatible with Thermo Scientific™ nanoViper™ Fingertight Fittings

Maximum Operating Pressure: 350 bar (5,000 PSI)

Part Number:
691-PO50-05VF
Wetted Material:
Stainless Steel, PEEK, Teflon, Fused Silica