The PrimeLine In-Line check valve incorporates many of the innovations of our HPLC self-priming replacement check valves. It is a rapid-closure cartridge valve that features very low resistance to flow, low internal volume, and operates independently of gravity. It features crush-proof construction. The valve may be used whenever flow needs to be restricted to one direction. Applications include prevention of post column derivatization fluids from back-flowing into the column in the event of mobile phase pump failure.
1/8" ball cartridge is protected from solvent contamination by a .250" diameter x .062" thickness 10 micron frit. If you see a large pressure drop across the valve with 1/8" ball cartridge version when you run at a very high flow rates, for example at 100 mL/min., please contact us for the solution.
This In-line check valve is High Temperature version that will operate at up to 200°C at 5,000 PSI.
Replacement cartridge for High Temperature version is not available.
Both 1/8" ball and 3/16" ball In-line CVs are suitable for SFC application.
In-Line 3/16" ball CV, P/N: 250-2001HT Specification
In-Line 1/8" ball CV, P/N: 250-2000HT Specification
Temporally 1/16" ball CV is not available due to the design improvement.